WARNING
Hazardous contact voltage
Risk of electric shock in case of incorrect connection.
For the electrical connection speci󹪝cations, refer to the information in Technical
speci󹪝cations (Page51).
At the mounting location of the device observe the applicable directives and laws for
installation of electrical power installations with rated voltages below 1000V.
WARNING
Missing PE/ground connection
Risk of electric shock.
Depending on the device version, connect the power supply as follows:
Power plug: Ensure that the used socket has a PE/ground conductor connection. Check that
the PE/ground conductor connection of the socket and power plug match each other.
Connecting terminals: Connect the terminals according to the terminal connection
diagram. First connect the PE/ground conductor.
WARNING
Unsuitable cables, cable glands and / plug-in connectors
Risk of electric shock and material damage
Only use cable glands/plug connectors that meet the requirements for electrical safety.
Tighten the cable glands according to the instructions in the respective manuals / operating
instructions.
Close unused cable openings for electrical connections.
When replacing cable glands, only use ones of the same type.
Make sure the cables are tight after installation.
Commissioning
5.2General information on commissioning
Series 6 and ULTRAMAT 23
Compact Operating Instructions, 01/2024, A5E45779144002-AB 45