Electrical characteristics
Analog outputs 1 analog current output per component, 0/2/4/NAMUR ... 20 mA, oating,
max. load 750 Ω
Relay outputs 8, with changeover contacts, freely selectable, e.g. for fault, loading ca‐
pacity 24V AC/DC/1A, oating, non-sparking
Binary inputs 3, dimensioned for 24 V, oating
• Pump
• AUTOCAL
• Synchronization
Serial interface ELAN (RS485), PROFIBUS-PA/DP as option
AUTOCAL function Automatic calibration (adjustment) of all channels with ambient air (in case
oxygen is present as measured component) or nitrogen (depending on
measured component),
adjustable cycle time from 1 … 24 hours. If 0 hours is entered, the cycle is
deactivated.
AUTOCAL cycle must be activated for certied emission measurements
Options Add-on electronics, with 8 additional digital inputs and 8additional relay
outputs, for e.g. triggering of automatic calibration, PROFIBUS PA/DP
Climatic conditions
Permissible ambient temperature
• During operation
• During transportation and storage
+5 ... +45 °C (+41 ... +113 °F) (limited when used with H
2
S sensor,
cf. Assembly/installation (Page25))
-20 °C ... + 60 °C (-4 ... +140 °F)
Permissible ambient humidity < 90% RH (relative humidity) during transportation and storage (Assembly/
installation (Page25))
Permissible ambient pressure 600 ... 1200 hPa (absolute)
Maximum operating altitude 2000m above sea level
Pollution degree 2 according to IEC 60664-1
Gas inlet conditions
Sample gas pressure
• Without pump
before device: On the inlet side, the overpressure of the gas with respect to
ambient pressure must not exceed 45hPa.
after device: On the outlet side, the gas outlet must be depressurized
against ambient pressure.
• With pump
before device: Depressurized suction mode
after device: On the outlet side, the gas outlet must be depressurized
against ambient pressure.
Sample gas ow 72...120l/h (1.2.. 2l/min)
Sample gas temperature 0...+50°C (+32...+122°F)
Sample gas humidity < 90%RH (relative humidity), non-condensing
Technical specications
7.1Technical specications ULTRAMAT 23
Series 6 and ULTRAMAT 23