P003948 Uscan User Manual Page | 9
Do not use the device or Power Supply if there are signs of damage.
Be aware of latex allergy. Some commercially available transducer covers contain latex. Refer to the
US FDA alert title: “Medical Alert: Allergic Reactions to Latex-Containing Medical Devices”, issued
March 29, 1991.
Check the connecting cable, connectors, and system housings before use for cracks or fraying. Do not
use if damaged.
No modification of this equipment is allowed.
This device contains no user-serviceable parts. Please contact Signostics customer support or your
Signostics distributor for maintenance or repair.
The internal probe battery is not to be replaced by the user – doing so may create fire or explosion
hazard.
Remove all particles and other matter from crevices and surfaces when cleaning the system and
components.
The device is supplied unsterile.
Clean and disinfect the transducer between patients.
Before cleaning or disinfection turn Uscan off and disconnect from the power supply.
Do not submerge the probe, the display or the power supply (charger) as electric shock could result.
The ultrasound probe is IPX1 which allows water falling vertically onto the probe. The display is
IPX0 and has no protection against ingress of water.
Clean and disinfect the system before placing in a bag for transport. Use the supplied Signostics probe
holder to store the probe. Clean and disinfect the probe holder regularly.
After cleaning or disinfection examine the ultrasound probe and display as appropriate for cracks or
leaks, and if damage exists discontinue use of the system and contact
Signostics customer support or
your Signostics distributor.
The user must not touch the any device connectors while in physical contact with the patient.
The Uscan probe is connected to the Uscan display running Signostics software to configure a
medical system. The display has been certified by Signostics as part of a medical system to IEC
60601-1: 3rd edition.
Do not connect the Uscan display to external computers or peripherals using the USB port unless the
system is outside the patient area. Failure to comply with these guidelines may result in electric shock.
Mounting the Uscan display on a mobile stand is configuring a medical system. Only use the
Signostics provided mobile stand accessory (P004013).
Only connect accessories that are specified as being compatible with Uscan. Contact Signostics
customer support or your Signostics distributor for information on compatible accessories and
systems.
If required, recharge the Uscan Probe only with the Signostics power supply (charger) provided
(P003198).
The Uscan power supplies are dedicated units to be used exclusively with the Uscan System only.
Do not open or modify the Signostics Power Supply P003198 or any other supplied power supplies –
Risk of electric shock
When using the optional Signostics mobile stand, if a Multiple Socket Outlet is required only use an
outlet and power cable certified to IEC 60601-1, and fix to the stand using a bolt and tool.
After storage at extreme temperatures, check the transducer surface temperature before applying to a
patient. A cold or hot surface may burn a patient.
Avoid musculoskeletal strain with prolonged use of Uscan.