P003948 Uscan User Manual Page | 35
Transducer cleaning and disinfection
Warning
Remove all particles and other matter from crevices and surfaces when cleaning the system and
components.
The device is supplied unsterile.
Clean and disinfect the transducer between patients
Before cleaning or disinfection turn Uscan off and disconnect from the power supply
Do not submerge the probe, the display unit or the power supply (charger) as electric shock could
result. The ultrasound probe is IPX1 which allows water falling vertically onto the probe. The display
may have no protection against ingress of water. Refer to the display user manual.
Clean and disinfect the system before placing in a bag for transport. Use the supplied Signostics probe
holder to store the probe. Clean and disinfect the probe holder regularly.
Caution
Use only recommended disinfection methods
Use abrasive cleaners, isopropyl alcohol or solvents sparingly, and if used immediately clean and
remove residual substances from the Uscan.
Do not heat sterilize any part of the Uscan device.
Cleaning
Signostics recommends the use of a soft cloth, lightly dampened with warm water or a mild soap solution for
cleaning the probe and transducer. Wipe surfaces with the moist cloth, and towel dry with a soft clean cloth.
For cleaning requirements of the display refer to its user manual.
Disinfection
The following disinfecting means can be used with the Uscan:
T-Spray II by Pharmaceutical Innovations of New Jersey
Sani-Cloth
®
Active by PDI of Flint, UK
Tuffie 5 Wipes by VernaCare of Lancashire, UK
Protex Disinfectant Wipes by Parker Laboratories of New Jersey
Any Isopropyl alcohol impregnated wipes with <= 70% w/w alcohol
Refer to Signostics website for additional disinfection options.
Caution
Minimize application of alcohol based disinfectant to colored overmold materials. Long term use may
result in material degradation. If alcohol based disinfectant is applied to the overmold, immediately
remove by wiping with a damp cloth.
The ultrasound probe and transducer may be disinfected using a wipe or spray method of disinfection. First
clean the ultrasound probe and transducer as above. Then wipe or spray the probe and transducer with
disinfectant, and leave for the disinfectant manufacturers recommended contact time. Remove any residue
with a soft cloth moistened with water. Do not allow any solutions to dry on the probe and transducer.
Warning
After cleaning or disinfection examine the ultrasound probe and display as appropriate for cracks or
leaks, and if damage exists discontinue use of the system and contact Signostics
customer support or
your Signostics distributor.