5 Commissioning
22 Minilab man-CH-PVF-C-V17-2
Commissioning 5
5.1 Adjustments
After the Minilab has been installed (See Section 4, “Unpacking and
Installation”) the following steps should be executed:
5.1.1 Compressed Air Pressure
1. Carefully open the Shut-off Ball Valve (Figure 8.2, Item 4) on
the rear of the device.
2. The operating pressure for the Minilab should be set to 4.5
bar with the Pressure Regulator (Figure 8.2, Item 3). This is
done by first pulling the rotary knob on the Pressure Regulator
outwards and then adjusting as necessary.
3. As the Piston (Figure 8.1, Item 3) descends, read the air
pressure setting on the Pressure Gauge (Figure 8.4, Item 1).
If the pressure on the Pressure Regulator is adjusted
without the squeezing Piston being moved, this can lead
to a false pressure display.
4. When necessary, the pressure should be reset during
operation of the Minilab (see Section 6 “Operating
Instructions”).
5. Push the pressure regulator rotary knob inwards to lock and
prevent unintentional pressure adjustment.
5.1.2 Power
Set the Main Switch (Figure 8.3, Item 1) to "ON". This turns on the
power for the Minilab control system and the display begins to
illuminate. After approximately 30 seconds the opening screen is
displayed.
During lengthy interruptions, e.g. overnight or on weekends, the Minil-
ab can be switched off with the Main Switch.