6 Operating Instructions
48 Minilab man-CH-PVF-C-V17-2
6.2.10 Wet Tensile Strength Measurement
1. Connect the Wet Tensile Strength Testing Apparatus (Model
PNZ), equipped with the Analog Output PNZ, to the interface
connection on the rear of the Minilab labelled, “Wet Tensile
Strength” (Figure 8.2, Item 2).
2. Select menu, “Wet Tensile Strength,” on the Operator Panel
(Figure 8.1, Item 6). The measurement results will automatically
be displayed on the Minilab.
3. Complete the wet tensile strength test with supplementary
equipment according to that equipment’s instructions &
standards.