Page 112 of 128
23. EMC compliance
The SLE1000 electro magnetic compatibility was tested to and meets the requirements of
the following relevant standards:
EN60601-1-2
EN61000-3-2
EN61000-3-3
See item 26 on page 21 for additional information.
Guidance and manufacturer's declaration - electromagnetic emissions
The SLE1000 is intended for use in the electromagnetic environment specified below. The
customer or the user of the SLE1000 should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment- guidance
RF emissions
CISPR 11
Group 1 The SLE1000 uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class A
The SLE1000 is suitable for use in all
establishments other than domestic, and may be
used in domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes, provided the following warning is
heeded:
Warning: This equipment/system is intended for
use by healthcare professionals only. This
equipment/ system may cause radio interference
or may disrupt the operation of nearby equipment.
It may be necessary to take mitigation measures,
such as re-orienting or relocating the SLE1000 or
shielding the location.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies