EasyManua.ls Logo

SLE SLE1000 - Power down

SLE SLE1000
128 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 30 of 128
Warning: If the unit is connected to Oxygen and the “O2 Calib Error alarm”
is triggered when entering either CPAP or Flow mode. Manually calibrate
the 100% oxygen point in either CPAP or Flow mode. If the “O2 Calib Error
alarm” is re-triggered remove the unit from service.
The unit will then display 21% oxygen and the user can now set
required percentage.
Note: In the first 4 hours the SLE1000 performs an adaptive calibration
routine. If the unit detects that the measured is lower than the set by 3%
then it forces a recalibration of the 100% oxygen. The unit will recheck the
difference every 8 minutes within the first four hours of operation.
Note: If the unit detects that the measured and set differ by ± 5% it
generates the “High Oxygen” or “Low Oxygen” alarm. In this instance the
user can manually recalibrate the 100% oxygen. See “100 % O
2
Calibration”
on page 40. This check is not limited to the first four hours of operation.
The unit automatically recalibrates the oxygen every 8 hours unless a calibration
error has occurred, (i.e. using the SLE1000 with air only) then the unit calibrates
every hour.
Note: If the SLE1000 is started using only air, the user can connect the
oxygen supply at any time, but will have to manually calibrate the oxygen
system.
Warning: If the oxygen supply fails after calibration whilst set to 21%
oxygen the unit will not alarm. The " No Oxygen Supply" alarm will only
become active when the user tries to set a percentage of oxygen above
21%.
7.3 Power down
Push and hold the power button on the rear of
the machine for 3 seconds. Re-press the
button to cancel the complete power fail
alarm.
Note: The unit can be turned on again after
3 seconds.

Table of Contents

Related product manuals