EasyManua.ls Logo

SLE SLE1000 - 17 Basic Operation; Monitoring; Operation

SLE SLE1000
128 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 88 of 128
17. Basic operation
The following chapter describes the basic operation and is not intended to be
prescriptive.
17.1 Monitoring
The recommended patient monitoring requirements are:
1. ECG/heart rate.
2. Respiration rate
3. Blood pressure (either by invasive or non invasive means).
4. Oxygen saturation.
5. Blood gas saturation by either transcutaneous carbon dioxide arterial /
capillary blood sampling.
6. Regular cranial ultrasound examinations (in the newborn).
7. Standard nursing care for Intensive Care patients.
17.2 Operation
Step 1 Setup the SLE1000 as described in section 14.4 on page 73.
Step 2 Ensure that the SLE1000 has been functionally tested as described in section 15.
on page 80.
Step 3 Ensure the unit has calibrated the oxygen system.
Note: If you are using the unit with air only, the “O2 CALIB ERROR” will be
triggered. Press the alarm button twice to clear the alarm message. The unit
will display “- -”. Do not set the % O2 above 21% as this will trigger the “NO
O2 SUPPLY” alarm.
Step 4 Set the SLE1000 into the required mode.
Step 5 Prepare the patient bonnet and select the prong or mask suitable for the patient.
Step 6 Attach the generator circuit to the patient.
Step 7 Set the Apnoea alarm time.
Step 8 Adjust the % oxygen, (If the unit is connected to air only leave the setting at 21%).
Step 9 Monitor the patient condition and adjust parameters accordingly.

Table of Contents

Related product manuals