Page 48 of 128
10.2 Audio paused
The user has the ability to mute the
alarm system.
When an alarm has been triggered, the
user pushes the audio paused button
once.
The unit will display the text “Audio
Paused”.
Note: The alarm message will
alternate between the alarm message
text and the audio paused text.
The mute will last for two minutes. The
alarm message will alternate with audio
paused message.
The user can cancel the mute by re-
pressing the audio paused button.
Note: Any latching alarms active
during audio paused mode are reset
on the second press of the button.
See “Latching and non latching
alarms” on page 98.
Note: The user needs to be aware that
the “Mains disconnect”, “No oxygen
supply” and “O2 Calib. Error” alarms
will not re-instate.
This is because the unit can operate
on battery power and on air only.