24. RS232
24.1 RS232 output specification:
Baud rate 19.2kbps
8-bit data
1 stop bit
No Parity
No hardware handshake
Continuously outputs current monitored values in an ascii text string that is 36 characters
long.
Characters (‘x’ denotes an ascii character):
0-15 'xxxxxxxxxxxxxxxx" Status in 16 characters
(“No Current Alarm” or alarm message)
16 ',' Comma separator
17-19 'xxx' BPM rate in 3 characters
20 ',' Comma separator
21-24 'xx.x' MAP in 4 characters
25 ',' Comma separator
26-29 'xx.x' Flow Rate in 4 characters
30 ',' Comma separator
31-33 'xxx' FiO
2
in 4 characters
34 0x0A Line feed
35 0x0D Carriage return
Note: the text “No Current Alarm” in the status field is only used for data output.