14.4.3 With air only from a medical air compressor
Step 7. Connect the air hose to
the SLE1000.
Step 8. Connect the air hose to
compressor.
Step 9. Turn on the compressor
and allow the pressure to
build up above 2 bar.
Note: The water trap / particulate
filter may bleed air due to the
auto drain mechanism not
closing. Block the outlet
momentarily to allow the
mechanism to close.
Step 10. Connect SLE1000 to
mains power. Do not turn
the SLE1000 on.
Proceed to “Functional
testing prior to patient
use” on page 80