OPERATING MANUAL PAT-1, PAT-2, PAT-2E, PAT-10 version 1.02
Press START/STOP button. The meas-
urement of resistance of protective con-
ductor will commence.
Correct result. The value
of leakage current is
below the set limit. Green
LEDs of PASS indication
are on.
Inorrect result. The value
of leakage current is
above the set limit. Red
LEDs of FAIL indication
are on.
The result is displayed on screen until ESC button is pressed, another test is run
by pressing START/STOP button, measurement mode is changed, settings mode
is activated or the test result is stored in the memory.
Additional information displayed by the tester
Measuring range is exceeded.
More messages displayed by the tester are described in section 2.1.2 of this manual.
Notes:
During the measurement in the test socket the mains voltage powering the tested device is
present.
During the measurement the tester automatically changes the polarity of the test socket and
repeats the measurement.
During the measurement of a faulty appliance, RCD switch may be triggered off.
- The tested device must be turned on.
- Ensure that the location of the tested device is isolated.
- The bandwidth of test current results from the measuring system with adjusted touch current which