S2/S2BW
Digital Color Doppler Ultrasound System
2.10 System Transportation
2.10.1 Moving the System
When moving or transporting the system, take the PRECAUTIONS listed below.
Before moving the system:
1. Completely power off the system.
2. Unplug the power cord.
3. Disconnect all cables attached to the system.
4. To prevent damaging the power cord, do not pull, stretch or bend it excessively.
5. Store the probes in their carrying cases or wrap them in soft cloth or foam to prevent damage.
6. Store gel and other essential accessories in the appropriate storage cases.
2.10.2 Transporting the System
Use extra care when transporting the system in a vehicle. After preparing the system as described above,
take the following additional PRECAUTIONS:
Before transporting, prepare and pack the system in the original package.
Secure the system firmly with straps (or as directed otherwise) within the vehicle to prevent excessive
vibrations during transport.
Drive carefully to prevent vibration damage. Avoid unpaved roads, excessive speeds.
After being transported, the system requires some time to reach thermal equilibrium with the environment.
Decrease or increase this time period (4 to 5 hours for 10
C difference) according to the initial temperature
difference of the system and the environment.
P/N: 4710.00149A01
2-13

Table of Contents

Other manuals for Sonoscape S2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sonoscape S2 and is the answer not in the manual?

Sonoscape S2 Specifications

General IconGeneral
BrandSonoscape
ModelS2
CategoryMedical Equipment
LanguageEnglish

Related product manuals