Rotors and accessories
16
Table 1: Rotors for Legend™ T 100V / 120V (differences of 230V instruments are shown in parentheses)
Rotor designation
Order no.
BIOshield
®
Rotor
4 x 250 ml
7500 6435
Highplate
®
Rotor
2 x 5 plates
7500 6444
Maximum permissible load [ g ] 4 x 600 2 x 500
Maximum speed nmax [ rpm ] 5350 5250
Maximum RCF value at nmax 5824 5083
Radius (max.) [ cm ] 18.2 16.5
Acceleration / deceleration time [ s ] 145 / 70 ( 85 / 70 ) 110 / 55 ( 85 / 60 )
Heating of samples at nmax
relative to room temperature 23°C,
Run time 1 hour
10°C 13°C
Aerosol-tight * yes yes
Autoclavable 121°C no
* Checked by TÜV (Technical Supervision Association) Product Service GmbH - Hamburg
!
"
"
"
!
#
$
%
#
&
'
$
(
)
%
&