Maintenance and care
51
Stopping the centrifuge
Stopping with preset run time
Normally the run time has been selected, and all you
have to do is wait until the centrifuge terminates the
run automatically at the end of the set time.
As soon as the speed reaches zero, the display
reads "End". You can now open the centrifuge by
pressing the "open lid" key
and remove your sam-
ples.
If the lid has not been fully lifted out of the lid lock,
the message "Lift lid" appears (you must manually lift
the lid).
You can manually stop the centrifuge at any time by
pressing the "Stop“ key
.
At this point the remaining run time is displayed.
Stopping with continuous operation
If you have chosen continuous operation, you must
stop the centrifuge manually. Press the "Stop" key
in the control panel. The centrifuge starts decel-
eration with the preset deceleration profile.
The display reads "End", and you can open the lid by
pressing the "Open lid" key
and remove your sam-
ples.
Temperature control during standby
Temperature control becomes active once the rotor
has been identified. This is the case after a centrifu-
gation run exceeding 300 rpm. At standby the display
reads "End".
If the rotor has not been identified (lid has been
closed and the "Start" key
has not yet been
pressed, speed panel shows "0" with flashing point),
the instrument regulates the temperature so that the
samples cannot freeze in any one of the usable ro-
tors.
If you find the systematic deviation of up to 4 K not
being sufficient, you have to start the rotor for a short
period until it is identified.
!
"
"
"
!
#
$
%
#
&
'
$
(
)
%
&