Operation
40
Aerosol-tight operation
Aerosol-tight rotors and tubes are
only to be opened in an approved
safety work bench when centrifuging
dangerous samples!
Paying attention to the maximum
permissible filling quantities is neces-
sary!
Correct operation when filling the sample tubes and
closing the rotor cover are prerequisites for aerosol
bio-containment.
Before use, the seals in the rotors and
rotor covers, as well as the aerosol-
tight caps, have to be checked for
abrasion or damage and slightly
greased.
Replace damaged O-rings and seals!
Use the special lubricant
7600 3500 only to grease the seals !
Spare parts are delivered with the rotor or may be
ordered separately.
Replace damaged or clouded caps
and lids of rotors and tubes imme-
diately.
The tubes are only to be filled such that the sample
does not reach the rim of the tube during centrifug-
ing.