Operation
41
BIOshield
®
-Rotor
Highplate
®
-Rotor
The aerosol-tight bio-containment of
air vessel rotors is only warrented in
a horizontal position!
If the air vessel rotors are filled with samples outside
the centrifuge (e.g. in a safety work bench), one has
to care, that during the installation into the centrifuge
the rotor should be kept in a vertical position.
Closing the aerosol-tight bio-containment
After greasing the seal, turn the cover until it sits
lightly on the bucket.
To achieve uniform seal, turn the cover clockwise by
1 ½ grip area ( approx. 15°). Use the setting marks
on the bucket as a guide.
Bucket mark
Grip area
!
"
"
"
!
#
$
%
#
&
'
$
(
)
%
&