Troubleshooting
63
Error troubleshooting
If problems other than those described in the following tables arise, you must contact
KENDRO service.
Error Symptom Possible causes and corrective measures
Displays remain
dark
The drive stops.
The rotor stops without
deceleration. The lid
cannot be opened.
Mains power failure or not connected
1. Is the main plug connected to the main socket?
2. Check the main connection.
3. If the main connection is OK, contact KENDRO service.
Displays fail
briefly.
The drive stops sud-
denly.
The rotor stops without
deceleration.
Brief interruption of main supply
1. Turn off main switch.
2. Check whether the mains power cord is connected properly.
3. Restart the centrifuge.
Error Symptom Possible causes and corrective measures