Troubleshooting
64
Lid cannot be
opened.
Pressing the "open lid" key
has no effect.
Lid not correctly engaged or lid warped.
1. Check whether main power is on and the instrument is switched
on (display is lit).
2. Press lid down in the middle of the front section once, and press
the "open lid" key.
3. If this is unsuccessful, you may open the lid using the emer-
gency lid release (see page 61).
Centrifuge is exceptionally
noisy.
Imbalance.
1. Stop the centrifuge by pressing the "stop" key, in case of emer-
gency, unplug mains power cord.
2. Wait until the centrifuge comes to a complete stop.
3. Check whether the rotor is properly loaded.
4. Check whether a broken tube, damage to the rotor or motor mal-
function is responsible for the noise.
If you cannot locate and solve the problem yourself, contact
KENDRO service.
!
"
"
"
!
#
$
%
#
&
'
$
(
)
%
&