Maintenance and care
57
• Use only disinfectants approved by KENDRO.
These agents are to be used according to the In-
structions for Use supplied with the respective
disinfectant:
− Aldasan 2000
− Carlitt Spray
− Coldspore
− Gigasept FF
− HBV Pump-Spray
− Incidin Liquid Spray
− Incidur Spray
− Incidin plus
− Kohrsolin iD
− Lysetol FF
− Lysoform
− Lysoformin 3000
− Sagromed
− Sagrotan
For all other disinfectants please consult KENDRO.
• You may disinfect the rotor and the accessories as
described in the following section. Be sure to fol-
low the pertinent safety procedures for handling in-
fectious material.
1. Unplug the main power cord.
2. Unscrew the collet chuck.
3. Grab the rotor with both hands and pull it perpen-
dicularly off the drive shaft.
4. Remove the centrifuge tubes and adapters, and
disinfect them or dispose of them as necessary.
5. Treat the rotor and the rotor cover according to the
instructions given for the disinfectant (soaking in
liquid or spraying). You must strictly observe the
specified reaction times!
6. Turn the rotor head down and drain it. Thoroughly
rinse rotor and cover with water.
7. Dispose of the disinfectant solution as according
to valid regulations.
8. Aluminum rotors have to be treated with anticorro-
sive grease subsequently
.