Troubleshooting
71
E-36
|
E-38
Rotor stops without brake
to standstill.
Instrument cannot be oper-
ated.
Overcurrent or error in current measurement.
Switch the instrument off and on again. If
the error persists, contact KENDRO service.
E-39
Rotor stops without brake
to standstill.
Instrument cannot be oper-
ated.
Speed control measurement exceeds permissible rotor speed.
Switch the instrument off and on again. If
the error persists, contact KENDRO service.
E-40
Rotor stops without brake
to standstill.
Instrument cannot be oper-
ated.
Acceleration of rotor too slow
Open the lid and check the load!
- Significant imbalance caused by missing bucket
- Tubes or adapter do not fit
(rotor is rubbing the chamber or the cover)
- Micro plates carriers on the motor cover
If the error persists, contact KENDRO service.