SS2000 Operator’s Manual
H
2
S in Natural Gas 27
3. Run conduit from the power source panel to the conduit hub on the
lower right side of the heated enclosure.
4. Pull ground, neutral and hot wires into the power terminal box inside
the heated enclosure.
5. Strip back the jacket and/or insulation of the wires just enough to
connect to the power terminal block.
6. For AC systems, attach the neutral and hot wires to the power
terminal by connecting the neutral wire to the terminal marked
“NEU,” the hot wire to the terminal marked “LINE,” as shown in
Figure 2–1.
Because the breaker or switch in the power panel will be the
primary means of disconnecting the power from the heater, the
power panel should be located in close proximity to the equipment
and within easy reach of the operator, or within 10 feet of the
analyzer.
An approved switch or circuit-breaker rated for 15 amps should
be used and clearly marked as the disconnecting device for the
heater.
ENCLOSURE
HEATER
NEU
LINE
Figure 2–3 AC connection terminal block for optional
enclosure heater.