Operating the SS2000 System
34 H
2
S in Natural Gas
Modes and Functions Defined
Use the keypad to access the following modes by pressing the key # key first
followed by a number (1, 2, 3, 4, 5, or 6) to activate a mode. The following
section explains each mode and the corresponding information that displays on
the LCD.
Mode 1: (Normal Mode)
Continuously displays updated measurements. Press the # key followed by the
1 key.
The measurements displayed are:
H2S — Concentration in the sample cell in units selected in Mode 2.
P — Pressure in the sample cell in units selected in Mode 2.
T — Temperature in the sample cell in units selected in Mode 2.
Note that every time another mode is accessed regardless of
whether a parameter has been changed, the analyzer requires
three to four minutes time upon returning to Mode 1 to re-
establish reference spectra before displaying a reading.
+
#
1
<NORMAL MODE>
H2S: 50 ppmv
P: 14.7PSI T: 76.1F
Figure 3–3 SS2000 measurement display.