SS2000 Operator’s Manual B1
Appendix B: Specifications
Table B-1 SS2000 H
2
S in Natural Gas Analyzer Specifications.
Performance
Concentration (H
2
S)
a
a. Consult factory for alternative ranges.
0–20 ppmv
Response Time
Display updates every 16 seconds
(software adjustable)
Repeatability ±0.5 ppmv
Application Data
Environmental Temperature Range –4 °F to 122 °F (–20 °C to 50 °C)
Environmental Relative Humidity 80% for temperatures up to 31 °C max
Altitude Up to 4000 m
Sample Inlet Pressure 25–30 psig (1.7–2.1 barg)
Cell Operating Pressure –4.5 to 10 psig (–0.3 to 0.7 barg)
Sample Flow Rate 13.9 SCFH [6.56 SLPM]
Contaminant Sensitivity
None for gas phase glycol, methanol, amines, or
mercaptans.
Electrical & Communications
Input Voltage
b
b. Supply voltage not to exceed ±10% of nominal. Transient over-voltages according to category II.
100–240 VAC, 50/60 Hz 1 (standard)
9–16 VDC or 18–32 VDC (optional)
Current (analyzer)
1A max @ 120VAC
0.5A max @ 240VAC
3.2 A @ 12 VDC, 1.6 A @ 24 VDC
Outputs
RS232 (all parameters)
4–20 mA loop (concentration only)
LCD Display Concentration, cell pressure, cell temperature
Physical Specifications
Size
66” H × 31” W × 15” D
(1676 mm H × 787 mm W × 381 mm D)
Weight Approx. 150lbs (68kg)
Pollution Degree 3
Sample Cell Construction 316L series polished stainless steel
Area Classification
Certification (Pending) CSA Class I, Div. 2, Groups C & D