To connect the sample inlet line:
1. Determine appropriate tubing route from the sample source to the
the analyzer.
2. Run stainless steel tubing from the customer-supplied upstream
regulator [set for 30 psig (2.1 barg)] to the sample inlet port of the
analyzer. Bend tubing using industrial grade benders, check tubing
fit to ensure proper seating between the tubing and fittings. Fully
ream all tubing ends. Blow out the lines for 10
15 seconds with
clean, dry nitrogen or air prior to making the connection.
3. Connect the inlet tube to the analyzer using the 1/4” stainless steel
compression-type fitting provided.