SS2000 Operator’s Manual
H
2
S in Natural Gas 3–11
2. Switch off the power to the analyzer using the switch or circuit-
breaker designated as the disconnection device for the equipment.
3. Disconnect the small in-line circular temperature/pressure
connector on the cable coming from the sample cell, as shown in
Figure 3–4.
4. Power up the analyzer by energizing the circuit to the analyzer.
Three to four minutes are required for the analyzer to establish
reference spectra before displaying a reading.
5. After establishing the reference spectra, the LCD displays Pressure
too High and the general alarm is triggered.
6. Depending on the 4/20 Alarm Option value, the current loop
output will either be 4 mA or 20 mA.
TEMP. &
PRESS.
CONNECTOR
Figure 3–4 Temperature and pressure sensor cable
connection.
Pressure too High
Dry P out of limit
Please Wait
Restart