Operating the SS2000 System
3ā6 H
2
S in Natural Gas
High Alarm Action
The High Alarm Action parameter determines the concentration threshold
above which the high alarm will be triggered. The value entered is compared
to the moving average over the integration time set by the Logger Rate. Enter
0 to turn off the high alarm or enter the threshold value desire followed by the
* key to set the high alarm and cycle to the next parameter.
Table 3ā1 Typical values for parameter setpoints.
Parameter Setting Function
Spectrum Average 1 - 100
Default = 64
Sets the number of
scans averaged for each
display reading
High Alarm Action 0 or >0 - Full Scale Sets the high alarm
threshold for moving
average
Logger Rate 1 - 300 readings
Default = 4
Sets display and current
loop integration time to
match data logging
interval
Temperature Unit 0 or 1 Sets the display unit for
temperature
Pressure Unit 0, 1, 2, or 3 Sets the display unit for
pressure
Concentration Unit 0 or 1 Sets the display unit for
concentration
Peak Tracking 0, 1, 2 Sets peak tracking capa-
bility to off, on, or reset
Cancel Scrub Alarm 0, 1 Turns off scrubber alarm
once activated
4/20 Alarm Option 0, 1, 2 Sets the current loop
state upon alarm condi-
tion
Cancel General Alarm 0, 1 Turns off general alarm
once activated
<SET PARAMETER MODE>
High Alarm Action
0
0-No Other-Alarm