SS2000 Operator’s Manual
H
2
S in Natural Gas 39
4/20 Alarm Option
The 4/20 Alarm Option determines the current loop state upon an alarm
condition. Enter 0 for no action, 1 for the current loop to assume a high state
upon an alarm condition, or 2 for the current loop to assume a low state upon
an alarm condition and press the * key to enter the value and cycle to the next
adjustable parameter.
Cancel General Alarm
The Cancel General Alarm parameter resets the general alarm once
activated. The general alarm is activated by system faults. Once activated, the
alarm will remain triggered until reset. Enter 1 followed by the * key to reset
the alarm.
This is the last parameter. Press the # key followed by the 1 key to return to
Mode 1.
Mode 4: (System Diagnostic Parameters - Channel A)
Displays system diagnostic data for Channel A. These values may be useful
when troubleshooting the system. Press the # key followed by the 4 key.
The diagnostic parameters displayed are:
DryTemp: Shows the temperature in the measurement cell when
scrubbed sample gas is flowing through it.
<SET PARAMETER MODE>
4/20 Alarm Option
0
0-No, 1-High, 2-Low
<SET PARAMETER MODE>
Cancel General Alarm
0
1–cancel
+
#
4
DryTemp:
DryPressure:
WetTemp:
WetPressure:
50.00C
1700.00mb
50.00C
1500.00mb