USER REFERENCE MANUAL
RoboSep™-S:
The Fully Automated
Cell Separator

Questions and Answers

  • M
    Mary OrtegaSep 9, 2025
    Why is sample dripping from pipette tips in RoboSep™-S Laboratory Equipment by STEMCELL?
    • M
      martinezrobertSep 9, 2025
      If you notice sample dripping from the pipette tips onto the carousel of your RoboSep™-S Laboratory Equipment by STEMCELL, it suggests a bad seal between the tip head and the disposable tip. Clean the tip head and replace the tip rack with new tips.
  • J
    Joseph MartinezSep 12, 2025
    Why cannot I add protocols in My Protocols tab of RoboSep™-S Laboratory Equipment by STEMCELL?
    • K
      Kristen ScottSep 12, 2025
      If you cannot add protocols in the My Protocols tab of your RoboSep™-S Laboratory Equipment by STEMCELL, it's likely because protocols are assigned to a quadrant. Remove any protocols assigned to a quadrant first, then add protocols to a user profile.
  • J
    joshua61Sep 13, 2025
    What to do if the robotic arm is stuck in RoboSep™-S Laboratory Equipment by STEMCELL?
    • T
      Tamara JonesSep 14, 2025
      If the robotic arm of your RoboSep™-S Laboratory Equipment by STEMCELL is stuck and bouncing in a fully extended position, the homing flag on the carousel may be misaligned. Power off the instrument and manually move the arm all the way down. Then, restart the instrument.
  • C
    Christopher MahoneySep 16, 2025
    Why tip rack separates when removing the lid on RoboSep™-S Laboratory Equipment by STEMCELL?
    • A
      Amy SmithSep 16, 2025
      If the tip rack separates when you remove the lid of your RoboSep™-S Laboratory Equipment by STEMCELL, it's likely due to excessive force. Make sure you are only removing the top lid.