00732467 Rev. D Page 35 of 110
8.0 Technical Specifications
8.1 General
Internal Design: Solid State
Output Configuration: Isolated (RF Floating)
Cooling: Natural Convection, Rear Panel Vent
Mounting: GoCart Mobile Cart or a flat, stable surface which
can withstand the weight of the gi4000 ESU.
Reference section 8.3 Dimensions and Weight.
Service Life: 7 Years
Equipotential Ground Connection: An equipotential ground connection is provided to
allow connection of the gi4000 ESU to a common
ground by way of a potential equalization connector.
This connection meets IEC 60601-1 requirements.
8.2 Classifications
Type of Protection Against
Electrical Shock: Class I (IEC 60601-1)
Degree of Protection Against
Electrical Shock: Type CF Equipment (IEC 60601-1)
Defibrillation Proof Applied Parts
Applied parts are the electrosurgical accessories
connected to the output receptacles of the unit,
including but not limited to: dispersive electrodes,
ArConnect® argon probe connector, ArC Smart™
argon coagulation probe(s), APC probes, monopolar
active cords, hot snares, hot forceps,
sphincterotomes, TouchSoft Coagulator® Monopolar
Probe, bipolar probes, irrigation tubing, etc.
Degree of Ingress Protection
(IP Rating): IPX0 (gi4000 ESU)
IP68 (gi4000 ESU footswitch)
Sterilization: The gi4000 ESU and accessories are NOT intended
to be sterilized or autoclaved.