8.11.4 Monopolar Pulse Blend Cut Mode
Figure 8-10
Figure 8-11
Figure 8-12
0
20
40
60
80
100
120
140
0 100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400 1500 1600 1700 1800 1900 2000
Output Power
Load Resistance
Pulse Blend CUT Mode
Power vs Load Resistance
Pulse Blend
CUT120
Pulse Blend
CUT 60
Pulse Blend
CUT Default
35
0
10
20
30
40
50
60
70
80
90
100
110
120
0 10 20 30 40 50 60 70 80 90 100 110 120
Output Power
Power Setting
Pulse Blend CUT Mode
Output Power vs. Power Setting into 500 ohm Load