00732467 Rev. D Page 77 of 110
12.0 Troubleshooting
12.1 On-Screen Notifications
Table 12-1
INFORMATION:
ACTIVE POWER
XX
No action required. gi4000 is delivering power XX.
No action required. gi4000 is irrigating with the
lavage pump.
INFORMATION: PURGING
Please wait.
Please wait for argon purge to complete.
INFORMATION:
Finish case, then
vent and replace.
Finish the procedure, and then vent and replace the
Argon Gas Bottle.
INFORMATION: VENTING
Please wait.
Please wait for venting to complete.
INFORMATION:
Bottle vented,
please replace.
It is now OK to replace the Argon Gas Bottle. See
section 9.4.3 System Installation and Setup:
Replacing Empty Argon Gas Bottle
WARNING:
Power Supply Calibration
Required
Finish procedure, then Contact STERIS Customer
Service for equipment service.
WARNING:
Main Calibration Required
Finish case, then Contact STERIS Customer
Service for equipment service.
Connect the dispersive electrode to the gi4000.
Verify positive connection of the patient dispersive
electrode connector in PSS Return Receptacle.
NOTE: When a Fault displays on the screen, reset the gi4000 ESU by turning the
Power ON/OFF Switch OFF and back ON. If the Fault clears the screen, the gi4000
can be put back into service. If the Fault does not clear from the screen, contact
STERIS Customer Service at 1-800-769-8226 or your local STERIS Endoscopy
Electrosurgery Specialist.