00732467 Rev. D Page 66 of 110
10.2 Procedural Use
10.2.1 Power Settings
When a Method/Mode has been selected, a default power setting is displayed that is
appropriate for any of the accessories pictured. A user can override this default for the current
application by pressing the up/down arrow keys shown on the power setting bar.
To change power using the UP/DOWN arrow keys on the power setting bar, touch the UP
arrow to increase the power in single Watt increments, or use the DOWN arrow to decrease
the power in single Watt increments.
The gi4000 ESU comes equipped with factory suggested DEFAULT SETTINGS. All power
defaults listed in Table 10-1 are suggested ranges for typical use denoted in watts (W). The
final power output chosen must always be the decision of the physician. Good electrosurgery
practice recommends using the lowest power possible for effective completion of the therapy.
Table 10-1
Method
Mode
Output
Power
Range
Default
Power
Tap-a-Tool™
Alternate
Default
Power
Argon
0 - 120
60
Argon Probe
45
Monopolar
TouchSoft®
Coag
0 - 120
50
TouchSoft
Coagulator
®
Probe
60
Forceps
40
Coag
0 - 120
20
Snare
25
Forceps
15
Blend Coag
0 - 120
20
Snare
25
Pulse Blend Cut
0 - 120
30
Snare
35
Sphincterotome
40
Blend Cut
0 - 200
30
Snare
35
Sphincterotome
40
Pulse Cut
0 - 120
35
Needle Knife
30
Sphincterotome
40
Cut
0 - 200
35
Needle Knife
30
Sphincterotome
40
Bipolar
0 - 60
20
Bipolar Probe
15