58
Message Explanation Action
Disc motor overloaded The disc motor was overloaded and
therefore overheated.
Wait till the motor has cooled off
and reduce the force before
restarting the process.
Sample motor overloaded The sample motor was overloaded
and therefore overheated.
Wait till the motor has cooled off
and reduce the force before
restarting the process.
Pump motor overloaded Water hose blocked or too much
swarf in coolant.
Replace coolant. Inspect pump.
Main supply voltage too low Supply voltage insufficient. Await normalisation of power grid.
Removal rate too low, phase time
exceeded
It has not been possible to
grind/polish the samples to the
expected level within 15 minutes.
Select a different surface or/and
increase the force.
The pressure in the compressed air
system is too low.
Sample holder unable to move,
upwards after process
It was not possible to move the
sample holder upwards after
finishing a process.
Force regulation error – contact
Struers service technician.
The pressure in the compressed air
system is too low.
Force regulation error It is not possible to reach the
selected force.
Force regulation error – contact
Struers service technician.
Air pressure too low Missing or incomplete air supply. Check that the air hoses are tight
and properly clamped.
If stock removal sensor (accessory)
is not installed this must be done
before the stock removal mode can
be used.
Stock removal sensor not installed Stock removal sensor has not been
installed or has not been calibrated.
If the stock removal sensor is
installed it is not calibrated
correctly, please look in the manual
for the stock removal sensor.