5.1.3 Installing the Documentation Station
Note All conduits should be provided with pull strings and should have minimum
bends or curves. All conduits have a maximum length of 45'. Conduits must have
insulated bushings on all open ends.
Install two (2) Ethernet connections for the SORN (Stryker's Remote Device Management System) and
SDC, respectively. A telephone connection may be desired for a nurse’s phone. Similarly, additional
Ethernet connections may be desired for a Nurses Computer and/or PACS Computer. If local Video
Conferencing via a CODEC is installed, an additional Ethernet connection is required. All additional
telephone/Ethernet connections will be specied by the Hospital.
ONE-BAY DOCUMENTATION STATION (DOC1)
US
Dimensions 25.5"W x 90"H x 30.2"D
Power ree (3) – 20 Amp circuits
• One (1) circuit for quad outlet in lower section.
• One (1) circuit for 2 duplex outlets in lower section.
• One (1) circuit for quad outlet in upper section.
• All documentation station circuits require critical power.
Data Per listed equipment
Backbox Per listed equipment
TWO-BAY DOCUMENTATION STATION (DOC2)
US
Dimensions 48"W x 90"H x 37"D
Power Recommend Six (6) - 20 amp circuits:
• One (1) circuit for quad outlet in lower section behind video router.
• One (1) circuit for quad outlet in lower section behind the light power sup-
ply box (if required).
• One (1) circuit for duplex outlet in lower section under touch panel.
• One (1) circuit for quad outlet in middle section behind digital capture de-
vice.
• One (1) circuit for quad outlet in upper le section (if required).
• One (1) circuit for quad outlet in upper right section (if required).
• All documentation station circuits require critical power.
Data Per listed equipment
Backbox Per listed equipment