29
S
8.1.4 Level System Cable Specications
• One (1) Belden 8723 or equivalent (maximum 1000')
• One (1) Belden 8241 or equivalent (maximum 1000')
ITEM-ITEM QTY SIZE
LWM – S
1
1"
8.2 Motorized Projection Screen (MPS)
Conduit One (1) 3/4" conduit
Back Box Junction box is located in le end cap of assembly
Power • Customer/Contractor to hardwire to standard electrical circuit prior to Stryker
installation
• Installation instructions included with assembly
Structural • Customer/Contractor to mount Stryker provided bracket to the wall in the
desired location with proper reinforcement to support the monitor
• Stryker Project Manager will provide mounting specications
8.3 Motorized Projection Screen Wall Switch (SWS)
Conduit One (1) 3/4" conduit
Back Box • 4"W X 4"H junction box with single-gang mud ring
• Flush mounted at standard light switch height
Power • Customer/Contractor to hardwire to standard electrical circuit prior to Stryker
installation
• Installation instructions included with assembly
ITEM-ITEM QTY SIZE
MPS – SWS
1
3/4"
8.4 Ceiling Mounted Projector
Cabling • Cables provided by the Hospital/Contractor for runs over 50'
• Cables pulled by the Hospital/Contractor
• Leave a 15' service loop at each end for terminations
Back Box • 4"W X 4"H junction box with single-gang mud ring
• Mounted 12" (minimum) behind projector mounting bracket
Power One (1) standard duplex outlet ceiling mounted adjacent to junction box
Structural • Customer/Contractor to mount Stryker provided bracket to the ceiling in the
designated location with proper reinforcement to support the projector prior
to Stryker installation
• Stryker Project Manager will provide mounting specications
8.4.1 SPI3 Cable Specications
• One (1) Belden 8723 or equivalent (maximum 1000')