ITEM-ITEM QTY SIZE
WRC/CRC – *
1
¾"
5.4.3 Wall Plates
Conduit One (1) ½" conduit
Back Box • 4"W x 4"H junction box with single-gang mud ring
• 4"W x 4"H junction box with dual-gang mud ring (Fiber optical DVI only)
• Mounted 18" above nished oor.
Power None required, but should be located next to outlet
ITEM-ITEM QTY SIZE
Copper DVI (CDP) – A
1
1 ½"
VGA/S-Video/BNC (TRP) - A
1 1"
Pass-rough (PTP) - A
1 1"
Fiber Optical DVI - A
1 1"
5.4.4 Observation Room Touch Panel, Microphone and Speaker (OBS)
LA
Conduit One (1) 1" conduit terminated
Cabling For runs over 50', Hospital/Contractor provides the following:
• Two (2) Belden 8723 / 88723 (maximum 1000')
• One (1) CAT6 cable (maximum 328')
• Cables require 15' service loop at both ends
• Cabling provided and pulled by Hospital/Contractor before Stryker installation
Back Box • 4"W x 4"H junction box with single-gang mud ring
• Mounted within 18" of touch panel location
Power One (1) standard outlet within 18" of touch panel location
ITEM-ITEM QTY SIZE
OBS – A
1
1"
5.4.5 Cable Run to Video Network Hub (HRN)
Conduit One (1) 2" conduit terminated at nearest corridor cable tray.
Cabling • Four (4) Belden 8241/88241 or equivalent (maximum 1000' - consult Project Man-
ager for runs over 1000')
• Two (2) Belden 8723 / 88723 or equivalent (maximum 1000' - consult Project Man-
ager for runs over 1000')
• One (1) CAT6 cable (maximum 328' - consult Project Manager for runs over 328'))
• Cables require 15' service loop at both ends
• Hub cabling provided and pulled by Hospital/Contractor before Stryker installation
ITEM-ITEM QTY SIZE
HRN – *
1
2"