7. Level Systems Room
Note All cabling must be provided and pulled by Hospital/Contractor prior to Stryker's
arrival.
Level System's are most oen installed in integrated doctor's oces, small conference rooms, and
pathology or nuclear medicine labs. is control system is placed in a cabinet next to or in the room's
cabinetry or desk. Cable passage must be allowed to the junction box for all equipment placed on the
desk or in the cabinetry.
7.1 Customer Supplied Casework (CSC)
Data Per listed equipment
Back Box Per listed equipment
Power One (1) - 20 amo circuit for outlets behind video router
Space
Requirements
Must allow for a minimum 2" cable passage between all components housed
inside.
Section housing video router must have an interior dimension of at least 21"W
X 16"H X 22"D
Section housing router must be vented.
Must allow for direct access to back boxes per requirements listed below.
Note Customer/Contractor responsible for installation of casework prior to Stryker
installation.
7.2 Casework Equipment
7.2.1 Level One System (S1)
LA
Dimensions 20.5"W X 8.5"H X 16"D
Data None
Back Box One (1) 6"W X 6"H X 4"D (minimum) junction box ush mounted at
same height as Level System
Terminate all integration conduits to this junction box
Power One (1) duplex outlet next to junction box
7.2.2 Level Two System (S2)
LA
Dimensions 20.5"W X 14"H X 19"D
Data None
Back Box One(1) 12"W X 12"H X 4"D (minimum) junction box ush mounted at
same height as Level System
Terminate all integration conduits to this junction box
Power One (1) duplex outlet next to junction box