8.11 Ceiling Mounted Microphone
Conduit One (1) 3/4" conduit
Cabling Cables provided by the Hospital/Contractor for runs over 50'
Cables pulled by the Hospital/Contractor
Leave a 15' (4.5m) service loop at each end for terminations
Cable Specications:
º One (1) Belden 8723 or equivalent (maximum 1000')
Back Box 4"W X 4"H junction box with single-gang mud ring
Flush mounted in desired location
Power None
ITEM-ITEM QTY SIZE
3CM – A (SPI3)
1
3/4"
ICM – ICW (IBC)
1
3/4"
8.12 Lutron Controller (LTC)
Conduit One (1) 1" conduit
Back Box Per Manufacturer
Power Per Manufacturer
Note e Stryker Innity Router is compatible with Grak Eye 3000/4000 Systems.
ITEM-ITEM QTY SIZE
LTC – A
1
1"
8.13 Lutron Lighting Control Integration Series
Conduit Cabling must run from the IBC to the Lutron Controller. Conduit determined by
Hospital/Contractor.
Cabling Cables provided by the Hospital/Contractor for runs over 50'.
Cables pulled by the Hospital/Contractor.
Leave a 15' (4.5m) service loop at each end for terminations.
Cable Specications:
º One (1) Belden 8723, 88723 or equivalent
Power None
8.14 Outside the OR Touch Panel, Keyboard, Mouse for SPI3
Conduit Cabling must run from the SPI3 to the Touch Panel, Keyboard, Mouse and/or
PC(s). One 2"(50mm) conduit is recommended. Conduit to be determined by
Hospital/Contractor.