27
S
7.2.3 Level System Touch Panel (T)
Dimensions Touch Panel: 11.9"W X 11.3"H X 6.6"D
Touch Panel Interface: 19"W X 2"H X 11"D
Power One (1) duplex outlet
7.3 Remote Level System Touch Panel (LTP)
Dimensions Touch Panel: 11.9"W X 11.3"H X 6.6"D
Touch Panel Interface: 19"W X 2"H X 11"D
Cabling Cables provided by the Hospital/Contractor for runs over 50'
Cables pulled by the Hospital/Contractor
Leave a 15' (4.5m) services loop at each end for terminations
Cable Specications:
º ree (3) Belden 8723 or equivalent (maximum 1000')
º One (1) Belden 8241 or equivalent (maximum 1000')
Conduit One (1) 1" conduit
Back Box 4"W X 4"H junction box with single-gang mud ring
Mounted within 18" of touch panel location
Power One (1) standard outlet within 18" of touch panel location
ITEM-ITEM QTY SIZE
LTP – S
1
1"