S
6.5.4 SD CODEC (L1)
Dimensions 19.5"W X 2"H X 12"D
Data One (1) Ethernet connection
Please consult Project Manager if ISDN is utilized.
Power One (1) standard outlet within 3'
6.6 IBC Podium with Touch Panel
LA
Note Customer/Contractor responsible for installation of podium prior to Stryker
installation.
Dimensions 32"W X 46"H X 24.3"D
Cabling Cables provided by the Hospital/Contractor for runs over 50'.
Cables pulled by the Hospital/Contractor.
Leave a 15' (4.5m) service loop at each end for terminations.
Cable Specications: Cables may vary depending on conference room congu-
rations. Please consult Project Manager for site specic requirements:
Touch Panel:
º One (1) Belden 8723 or equivalent (maximum 1000')
º One (1) Extron MHR5 or equivalent (maximum 150')
Laptop Connection:
º One (1) Belden 8723 or equivalent (maximum 1000')
º One (1) Extron MHR5 or equivalent (maximum 150')
PC Connection:
º One (1) Belden 8723 or equivalent (maximum 1000')
º One (1) Extron MHR5 or equivalent (maximum 150')
Podium Microphone
º One (1) Belden 8723 or equivalent (maximum 1000')
Network
º Two (2) CAT6 cables (maximum 328')
Conduit Two (2) 2" conduits
Back Box 12"W X 12"H junction box
Flush mounted in the oor under the podium
Power One (1) standard outlet
Flush mounted in the oor beneath podium adjacent or within 12" X 12" junc-
tion box
ITEM-ITEM QTY SIZE
IPD – ICW
2
2"