CClleeaanniinngg aanndd ddiissiinnffeeccttiinngg wwiitthh wwiippeess
For United States only. Confirm availability for your configuration or region. Call Stryker Customer Service: 1-800-327-
0770.
Stryker’s preferred wipes (2060-000-001 6'' x 10'' or 2060-000-002 9'' x 12'') include the following active ingredients:
n-Alkyl (60% C14, 30% C16, 5% C12, 5% C18) dimethyl benzyl ammonium chloride - 0.154%
n-Alkyl (68% C12, 32% C14) dimethyl ethylbenzyl ammonium chloride - 0.154%
Isopropanol - 21.000%
Non-active ingredient: Ethylene Glycol Monobutyl Ether < 3%
NNoottee -- For safety information, read the product label.
To clean or disinfect the external product surface:
1. To clean, wipe external surfaces with a fresh, clean wipe to remove all visible soils. Repeat as necessary until the
product is clean.
NNoottee
Use as many wipes as necessary.
Complete step 1 before you disinfect.
2. To disinfect, wipe external surfaces with a fresh, clean wipe until wet. Allow the external surface to remain wet for two
minutes at room temperature.
3. Allow the product to dry before you return it to service.
EN 42 3009-009-001 Rev B.0

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker ProCuity Series and is the answer not in the manual?

Stryker ProCuity Series Specifications

General IconGeneral
Bed TypeHospital Bed
Weight Capacity500 lbs (227 kg)
Integrated ScaleYes
Bed Exit SystemYes
Side Rail ControlsYes
CategoryMedical Equipment
Monitoring ParametersWeight, bed exit
DisplayTouchscreen
AlarmsBed Exit, Low Battery

Related product manuals