2. On the SSccaallee screen, press the AAdddd EEqquuiipp.. button (E) (
Footboard control panel - Scale
(page 31)).
3. One item at a time, add or remove the desired equipment from the product.
NNoottee -- Do not touch the product while the product weighs the equipment.
4. Press the SSaavvee button to save the current weight and equipment count.
FFoooottbbooaarrdd ccoonnttrrooll ppaanneell -- BBeedd eexxiitt
The BBeedd EExxiitt screen displays the bed exit functions of the product.
A Home
Returns to the HHoommee screen (
Footboard control
panel - Home
(page 29))
B Low
Allows the patient to move freely on the bed. Alarms
when the patient moves 50 percent body weight out
of the designated zone.
C Medium
Allows for limited movement. Alarms when the
patient approaches the siderail or the foot end of the
bed.
D
High
Allows minimal movement. Alarms when the patient
moves out of the tightly restricted zone.
E Turn off Turns off bed exit
AArrmmiinngg oorr ddiissaarrmmiinngg bbeedd eexxiitt
WWAARRNNIINNGG -- Do not use bed exit to replace patient monitoring protocol. Bed exit is intended only to aid in the detection of a
patient exiting the product.
When armed, bed exit monitors the patient’s position on the product.
To arm bed exit:
1. Set the scale to zero/tare if not already performed. See
Zeroing/taring the scale
(page 32)
NNoottee -- If you do not set the scale to zero before you place a patient on the product, bed exit may not operate as
intended.
2. Position the patient on the product.
EN 34 3009-009-001 Rev B.0