Calibration summary screen
The calibration summary screen displays the following information:
Correction: Value of the current correction factor. This will be 0.000 for
factory calibration and will be updated when a user calibration is
performed.
Calibration type: Factory calibrated or user calibrated.
Last calibrated: Displays the date of the previous user calibration.
Next calibration required: Displays the date when the next user calibration is due, depending on
the frequency reminder in the Settings.
❖ Touch ‘RECALIBRATE’ to continue and recalibrate the instrument.
Note: One, two or three certified melting point determinations can be used for recalibration. It is
recommended that the standards used bracket the expected melting point range of your samples.
❖ Touch ‘SELECT’ to search for the result file of the first standard to be used.