EasyManua.ls Logo

Suez CheckPoint Pharma - Page 54

Suez CheckPoint Pharma
112 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CheckPoint Pharma and CheckPoint
e
Operation and Maintenance Manual
SUEZ © 2018 54 of 112 DLM 97200-03 EN Rev. A
3. Route the I/O cables into the Sensor through the power module panel pass-through port. See
Figure 13: Routing the I/O Cables (Detail).
4. Route the cables through the first grommet and then through the second grommet, to the I/O
terminal blocks (see Figure 12: Routing the I/O Cables). If needed, you can cut a notch in the
grommets to make wire manipulation easier.
Figure 12: Routing the I/O Cables
Figure 13: Routing the I/O Cables (Detail)
5. For 4-20 mA, binary input, and alarm wiring, connect the wiring to the appropriate pins. Consult
the tables below for pin functionality on each terminal block.
Wire should be 22-12 AWG, rated to 300 Volts. Strip length should be 8-9 mm (.33 in). To attach
the output connections, first remove the terminal block from the I/O board by firmly grasping the

Table of Contents