1. Specifications
Model
!
EC-4110
Measuring modes
!
Resistivity/Conductivity/Temp.
Resistivity
0.00 M·cm~20.00 M·cm
Conductivity
0.00 S/cm~200.0 mS/cm manual or auto range selectable
Ranges
!
Temp.
!
-30.0~130.0˚C
Resistivity
0.01 M·cm
Conductivity
0.01 S/cm
Resolutions
!
Temp.
!
0.1˚C
Resistivity
±1% ± 1Digit
Conductivity
±1% ± 1Digit
Accuracy
!
Temp.
!
±0.2˚C± 1Digit
Temperature
Compensation
!
NTC30K or PT1000 or
Manual temperature compensation selectable
Calibration mode
!
(1)Cell constant adjustment (2)Standard solution calibration
Ambient Temp.
!
0~50 ˚C
Storage Temp.
!
-10~70 ˚C
Cell Constant
!
0.01, 0.05, 0.1, 0.5, 10.00 cm
-1
fixed,
freely selectable 0.008~19.99 cm
-1
Temperature Coefficient
!
Linear temperature compensation from (0.00%~ 40.00%)
and Non-Linear compensation
Display
!
Large LCD display with environment light sensor
auto/manual illumination function
Analog output 1
!
Isolated DC 0/4~20mA corresponding to main measurement,
max. load 500:
Analog output 2
!
Isolated DC 0/4~20mA corresponding to Temp., max. load 500:
Contact
!
RELAY contactΔ240VAC 0.5A Max.(recommend)
Settings
!
Activate
!
Two sets of individual HIGH or LOW programmable control
Wash
RELAY contact: ON 0~99min. 59sec. / OFF 0~999hr 59min.
Certification
IP65 (NEMA 4)
Power Supply
100V~240VAC±10%Δ50/60HzΔ5W max.
Installation
Wall or Pipe or Panel Mounting
Dimensions
96m × 96mm × 132mm (HuWuD)
Cut off Dimensions
93 mm u 93 mm (HuW)
Weight
0.5Kg
Note: The specifications are subject to change without notice.