SV 104A USER'S MANUAL_______________________________________________________
36
the calibrator is not properly attached to the instrument
there are external acoustic disturbances such as high noise levels close by
the calibrator or the measurement channel (for example the microphone) is damaged
Note: During the calibration measurement, the external disturbances (acoustic noise
or vibrations) should not exceed a value of 100 dB (when using a calibrator that generates a level
of 114 dB).
8. Press the <ENTER> key in order to accept the measurement result.
Figure 4-11 Calibration confirmation screen
Note: If a calibration factor does not meet tolerance criteria ±2 dB, you still can manually
accept the microphone, but the results could possibly be affected - see Figure 4-12.
Figure 4-12 Calibration - microphone not in tolerance screen
9. Post calibration. If enabled, the post processing is performed automatically under acceptance
of calibration measurement.
Note: If it is required to perform so called post-calibration of the instrument. In case the Post
Calibration is pre-programmed within configuration setup the SV 104A automatically adds the results to
the previously saved files.
new
calibration factor
previous

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Svantek SV 104A and is the answer not in the manual?

Svantek SV 104A Specifications

General IconGeneral
BrandSvantek
ModelSV 104A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals