• the calibrator is not properly attached to the instrument
• there are external acoustic disturbances such as high noise levels close by
• the calibrator or the measurement channel (for example the microphone) is damaged
Note: During the calibration measurement, the external disturbances (acoustic noise
or vibrations) should not exceed a value of 100 dB (when using a calibrator that generates a level
of 114 dB).
8. Press the <ENTER> key in order to accept the measurement result.
Figure 4-11 Calibration confirmation screen
Note: If a calibration factor does not meet tolerance criteria ±2 dB, you still can manually
accept the microphone, but the results could possibly be affected - see Figure 4-12.
Figure 4-12 Calibration - microphone not in tolerance screen
9. Post calibration. If enabled, the post processing is performed automatically under acceptance
of calibration measurement.
Note: If it is required to perform so called post-calibration of the instrument. In case the Post
Calibration is pre-programmed within configuration setup the SV 104A automatically adds the results to
the previously saved files.