EasyManuals Logo

Svantek SV 104A User Manual

Svantek SV 104A
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #62 background imageLoading...
Page #62 background image
SV 104A USER'S MANUAL_______________________________________________________
62
6.5.7.1 Calibration settings panel
Sometimes it is required to perform so called post-calibration of the instrument. The position Post Cal.
enables the user to perform additional calibration after a measurement session and add the results to the file
saved in the memory. In the panel below, there are three options for saving results: not to save “Off”, save in
the last file “Last File” or save in the files which will be created after last calibration (“Files after last
calibration”). Auto-calibration can be disabled if required for any reason.
Figure 6-14 Calibration settings panel
6.5.7.2 Statistical levels settings panel
In the Statistical Level panel, it is possible to define ten percentile statistical levels, named from N1 to
N10. The default statistical levels have the following settings: 1, 10, 20, 30, 40, 50, 60, 70, 80 and 90. All values
have to be within the integer range [1, 99]. Each value can be set independently from the others.
Figure 6-15 Statistical levels settings panel

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Svantek SV 104A and is the answer not in the manual?

Svantek SV 104A Specifications

General IconGeneral
BrandSvantek
ModelSV 104A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals