Why does my TA Instruments Chiller compressor shut off and on?
- DDerrick FoxAug 6, 2025
If the compressor in your TA Instruments Chiller shuts off and restarts every 3-4 minutes, possible causes include low voltage, high ambient temperature, dirty condenser fins, or a malfunctioning fan. Check the voltage under load against the serial tag specifications (it must be within 5%). Ensure the room temperature meets the installation requirements. Clean the condenser fins. Verify the fan's operation by feeling for airflow across the condenser; if the fan isn't working, contact TA Instruments.